Detector

VWD and diode array (DAD/PDA)

A variable wavelength detector picks one wavelength before the cell. A diode array sends all the light through the cell, then splits it onto hundreds of diodes, so you get the full spectrum of every peak.

Both are UV detectors. The difference is where the wavelength is chosen.

In a variable wavelength detector (VWD) the grating sits before the flow cell. Only the wavelength you set goes through the sample, and one photodiode reads it. Agilent calls this direct optics.

In a diode array (DAD, or PDA at Waters) the whole lamp spectrum goes through the cell first. After the cell a grating spreads the light onto an array of diodes, 1024 on Agilent’s DAD WR and 512 on the Waters ACQUITY PDA. So every data point is a full spectrum. Agilent calls this reversed optics.

What the spectrum gives you: you can pick the wavelength after the run, compare the spectrum of a peak with a standard, and check whether a peak is pure. The VWD gives a quieter single trace at one wavelength. Agilent’s 1260 VWD is specified below ±0.25 × 10⁻⁵ AU noise.

The student page covers spectra and peak purity.

The numbers

VWD optics (Agilent)
Direct optics: the chosen wavelength passes the cell, one photodiode reads it 1
DAD optics (Agilent)
Reversed optics: all wavelengths pass the cell, a grating splits them onto a diode array 1
UV range
Compounds must absorb between 190 and 600 nm 1
Diodes (Agilent DAD WR)
1024-element photodiode array 2
Diodes (Waters ACQUITY PDA)
512 photodiodes, 1.2 nm optical resolution 4
Noise (Agilent 1260 VWD)
Below ±0.25 × 10⁻⁵ AU at 230 nm, single wavelength 2
Linear to (Agilent 1260 VWD / DAD WR)
Above 2.5 AU / above 2 AU (5%) 2
VWD with two wavelengths (Agilent G7114A)
Data rate drops from 120 Hz to 2.5 Hz 2

Same part, different names

Agilent
VWD (variable wavelength), DAD (diode array), MWD (multiple wavelength) 2
Waters
PDA (photodiode array), eλ PDA 3

What goes wrong

  • Peaks distorted when a VWD is set to two wavelengths in a fast runCheck: On Agilent's 1260 VWD, dual wavelength mode drops the data rate to 2.5 Hz. Use one wavelength, or a DAD. 2
  • Negative peaksCheck: The sample absorbs less than the mobile phase at that wavelength (Dolan). Change the wavelength, or use a mobile phase that absorbs less. 5

Check yourself

Tap an answer to see if you are right and why.

In a diode array detector, where is the light split into wavelengths?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, The LC Handbook. Detector, the organ of perception, pp. 24-25.
  2. Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 36 (VWD G7114A), p. 52; Table 38 (DAD WR G7115A), p. 55.
  3. Waters, ACQUITY UPLC PDA and eλ PDA Detectors: Overview and Maintenance Information (715002209). Detector overview, chapter 1.
  4. Waters, ACQUITY UPLC PDA Detector Getting Started Guide (71500108703). Detector optics, p. 1-2.
  5. ACE / Advanced Chromatography Technologies, HPLC Troubleshooting Guide (J.W. Dolan, for ACE). Negative peaks, section 3.3.