Detector

Flow cell

The small window the mobile phase flows through while the light passes across it. Path length sets the signal; volume sets how much the peaks spread.

The flow cell is where the measurement happens. The light crosses a short channel, and the peak passes through it.

Two numbers matter. Path length sets the signal: Beer-Lambert says absorbance goes up with the length of the light path, so a 10 mm cell gives more signal than a 3 mm one. Volume sets how much the peak spreads inside the cell. Waters’ rule: keep the cell at one-tenth of the peak volume or less.

The trouble is that the two pull against each other. On older designs a smaller cell meant a shorter path and less signal. Agilent’s standard 1200 cell is 10 mm and 13 µL; the micro cell is 3 mm and 2 µL. Newer light-guiding cells (Waters’ Teflon AF cell, Agilent’s Max-Light) keep a long path in a small volume.

One number people forget: the pressure rating. Flow cells are built for low pressure. Agilent’s standard VWD cell is rated 40 bar, the DAD cell 120 bar, the Waters PDA cell 69 bar. Anything that adds backpressure after the cell counts against it.

The numbers

Beer-Lambert
Absorbance is proportional to concentration times path length 1
Standard cell (Agilent 1260 DAD WR)
10 mm path, 13 µL, 120 bar maximum 2
Semi-micro / micro (Agilent)
6 mm, 5 µL / 3 mm, 2 µL 2
Standard cell (Agilent 1260 VWD)
10 mm path, 14 µL, only 40 bar maximum 2
Waters ACQUITY PDA cell
69 bar (1000 psi) maximum 5
Size rule (Waters)
Cell volume 1/10 of the peak volume or less 4
Size rule (Agilent)
Tubing, injection and detector volume together under 10% of the column volume 1

Same part, different names

Agilent
Standard, semi-micro, micro, Max-Light cartridge cell 2
Waters
Light-guiding flow cell (analytical, high sensitivity) 4

What goes wrong

  • Peaks broader on a short, narrow columnCheck: Standard cell on a small column. Agilent recommends the semi-micro or micro cell for high-efficiency columns, at some cost in sensitivity. 1
  • Leak at the cell, or intensity dropsCheck: Agilent's reasons to clean or replace the cell. A cracked cell window gives drift too (Shimadzu): replace the window. 3
  • Cell leaks after a restrictor, a second detector or a long waste line was added after itCheck: Too much pressure on the cell. Waters: whatever sits after the PDA cell must give at least 17 bar and no more than 69 bar. Check your cell's own rating. 4

Check yourself

Tap an answer to see if you are right and why.

Which flow cell gives the larger UV signal for the same sample?

Go deeper

The more detailed pages (marked Students) are for students of HPLC Operation and Troubleshooting, a live group. How to become a student. Students: sign in with your email when asked.

Sources

  1. Agilent Technologies, The LC Handbook. Lambert-Beer, p. 24; detector sensitivity, p. 26; extra-column volume, p. 42; flow cells for Poroshell columns, p. 89.
  2. Agilent Technologies, 2016, InfinityLab LC Series Specification Compendium (01200-90062). Table 36 (VWD), p. 53; Table 38 (DAD WR), p. 56.
  3. Agilent Technologies, InfinityLab LC Series Diode Array Detectors User Manual, maintenance chapter. Overview of maintenance, p. 161.
  4. Waters, ACQUITY UPLC PDA and eλ PDA Detectors: Overview and Maintenance Information (715002209). Operating principles of the light-guiding flow cell, p. 5; clearing bubbles from the flow cell, p. 56.
  5. Waters, ACQUITY UPLC PDA Detector Getting Started Guide (71500108703). Flushing the flow cell, p. 5-7.
  6. Shimadzu, HPLC Troubleshooting Guide. Baseline drift, p. 17.