Detector
Outlet restrictor (backpressure regulator)
A small regulator or a length of narrow tubing on the detector outlet. It keeps a little pressure on the flow cell, so gas stays dissolved and does not make bubbles and spikes.
Gas stays dissolved under pressure. In the pump and column the pressure is high. In the flow cell, after the column, it is close to nothing, and the gas comes out as small bubbles in the light path. You see them as spikes.
The fix is to keep a little pressure on the cell. Waters fits a regulator that holds at least 17 bar after the PDA cell, and says it is required when the PDA is the last detector. If a mass spectrometer follows the PDA, the tubing to it does the same job.
Without a regulator, a length of narrow tubing works. Thermo: 30 to 90 cm of 0.23 mm ID on the detector outlet, which gives about 30 to 50 psi at 1 mL/min for 90 cm.
There is an upper limit too. The cell has a pressure rating (69 bar on the Waters PDA cell), so don’t stack restrictors.
The numbers
- Why (Waters)
- Pressure after the column is much lower, so dissolved gas comes out in the cell and gives spikes 1
- Waters ACQUITY PDA regulator
- Keeps at least 17 bar (250 psi) after the cell; required when the PDA is the last detector 1
- Upper limit (Waters)
- No more than 69 bar (1000 psi) on the PDA cell 1
- Tubing instead (Thermo)
- 30 to 90 cm of 0.23 mm ID tubing on the detector outlet 2
- What that gives (Thermo)
- 90 cm: 30 to 50 psi at 1 mL/min 2
- Waters PDA tubing
- 30 to 60 cm of 0.23 mm (0.009 in) ID on the waste outlet 1
Same part, different names
What goes wrong
- Sharp spikes in the baselineCheck: Bubbles in the cell. Degas the mobile phase and put a restrictor on the detector outlet (Agilent, Dolan). 3
- Spikes started after the detector was moved or the waste line changedCheck: Check the regulator or the narrow outlet tubing is still there. Waters needs it when the PDA is the last detector. 1
Check yourself
Tap an answer to see if you are right and why.
Sources
- Waters, ACQUITY UPLC PDA and eλ PDA Detectors: Overview and Maintenance Information (715002209). Plumbing the detector, p. 21; clearing bubbles from the flow cell, p. 56; troubleshooting, p. 79.
- Thermo Fisher Scientific, Successful HPLC Operation: A Troubleshooting Guide. Section 3.3.1, air bubble trapped in detector flow cell.
- Agilent Technologies, The LC Handbook. Troubleshooting, baselines, p. 128.
- ACE / Advanced Chromatography Technologies, HPLC Troubleshooting Guide (J.W. Dolan, for ACE). Ghost peaks and spikes, section 3.