Methods and method sets

Method set

A method set points to the instrument method and, if you want, the processing, report and export methods. With several channels, 3D data or derived channels, you must process with it.

Think of the method set as a recipe card that points to the other methods. It is what Empower uses to run the instruments, and it can also tell Empower how to process, report and export afterwards.

When you need it for processing

Waters’ tip walks through the cases:

Data Process with
One 2D channel (e.g. UV at 280 nm) Any of the three: acquisition method set, a specified method set, or the processing method. Same results.
Several 2D channels (UV plus pump pressure) A method set whose channel table lists only the channels to process, each with its processing method
Two UV channels, or UV plus RI A method set, one processing method per channel
3D data (PDA, fluorescence, QDa) A method set, including any derived (extracted) channels

Different processing for different injections

If some injections need one processing method and others another, Waters’ answer is to use two method sets in the sample set method, each with its own processing method, and then process with the acquisition method set.

What a reviewer checks

  • Which method set and processing method made the result (both are recorded).
  • On PDA data: that the result came from a method set, with the right extracted wavelength.
  • That a diagnostic channel was not processed and reported by mistake.

The numbers

What it holds
A group of methods, e.g. instrument, processing, report and export 1
Minimum for an injection
A method set with an instrument method 1
Must process with a method set
Multiple channels per injection, 3D data, or derived channels 1
One 2D channel
Acquisition method set, a specified method set, or the processing method all give the same results 2
Channel table
Pairs each channel with its own processing method 2

What goes wrong

  • PDA data processed with only a processing method gives wrong or missing peaksCheck: Waters: with 3D data, several channels or derived channels, process with a method set. 1
  • The pressure channel gets integrated with the UV channelCheck: Waters: list only the UV channel and its processing method in the method set's channel table, so the diagnostic channel is left alone. 2
  • The next run processed with an old processing methodCheck: The method set points to a processing method. After you change the processing method, check the method set still points where you meant. 2

Check yourself

Tap an answer to see if you are right and why.

You process PDA (3D) data. What must you process with?

Sources

  1. Waters, 2017, Empower 3 Getting Started Guide (715005480). Key concepts, pp. 7, 9-10; processing the sample set, p. 102.
  2. Waters Help Center, Empower Tip #330: When to process data with a processing method versus a method set. Processing method versus method set.