Processing and integration
Int Type: the two letters on every peak
Each peak gets two letters that describe its start and end. Capitals mean the processing method did it. Lowercase means someone moved it by hand.
The Int Type column in the Peaks table is the fastest review tool in Empower. Two letters per peak: how it starts, how it ends.
The letters
| Letter | Boundary |
|---|---|
| B | Baseline |
| V | Valley (drop line between touching peaks) |
| S | Shoulder (a valley worked out from the curve) |
| R | Round |
| G | Gaussian skim |
| T | Tangential skim |
| X | Crossover between a positive and a negative peak |
So BB is a clean, resolved peak. BV, VV, VB are the first, middle and last peaks of a cluster.
Capitals and lowercase
Capitals mean the processing method did it. Lowercase means a person moved it. Bb means the start was automatic and the end was moved by hand. bb means both ends were moved.
Using it for review
Waters’ tips on exception-based review:
- In the project window, Peaks tab, make a View Filter that lists peaks with lowercase letters. Now you see every manual integration in the project at once.
- Make a Boolean peak custom field that returns “Manual Integration” when any letter is lowercase.
- Make a result custom field from it, so a result says whether any of its peaks were integrated by hand. Put that field on the report.
Then the reviewer starts with those results instead of reading every peak.
The numbers
- BB
- Baseline-resolved peak 1
- BV / VV / VB
- Starts a cluster / inside a cluster / ends a cluster 1
- S, R
- Shoulder, round (need Detect Shoulders) 1
- G, T
- Gaussian skim, tangential skim (need the event) 1
- X
- Crossover between a positive and a negative peak 1
- Lowercase
- Manual integration, e.g. Bb = end moved by hand 1
- Find them all
- View Filter on the Peaks tab for lowercase letters 2
- Flag them in reports
- Boolean peak custom field, then a result custom field 3
What goes wrong
- You need every manually integrated peak in a projectCheck: Waters: make a View Filter on the Peaks tab for lowercase Int Type letters. 2
- Manual integration is invisible on the printed reportCheck: Waters: a Boolean peak custom field returns 'Manual Integration' when any letter is lowercase; a result custom field carries it up to the result and the report. 3
Check yourself
Tap an answer to see if you are right and why.
Sources
- Waters, 2017, Empower 3 Data Acquisition and Processing Theory Guide (715005481). Section 2.1.5, Table 2-1, pp. 14-15.
- Waters Help Center, Empower Tip #291: identifying manually integrated peaks in results. Steps 1-4.
- Waters Help Center, Empower Tip #363: How to easily see that a peak has been manually integrated. Boolean custom field.
New modules by email
I am adding the Empower pages area by area, together with the live group Empower 3 for Beginners. Leave your email and I will send you each one when it is ready.
My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy