nalam.ca Learn
Sign in Join a group

Point at a module, then click to open it. The blue tubing opens Tubing and fittings.

Tubing and fittings

Capillary tubing ID and colour bands

1/16" OD tubing after the pump. The inside diameter decides how much pressure and volume it adds.

After the pump, the tubing is 1/16 inch outside diameter, stainless steel or PEEK. The outside is the same on every piece. The inside diameter is what changes, and you can’t see it by eye.

IDEX puts a colour band on its precut stainless steel tubing to show the ID. This is IDEX’s own scheme. Other makers may use different colours, so check the maker’s sheet before you trust a colour.

The volume grows with the square of the ID. Blue 0.010 inch tubing holds 4 times the volume of red 0.005 inch per cm. 10 cm of yellow 0.020 inch holds about 20 µL.

The numbers

Red band
0.005" ID (0.127 mm). About 0.13 µL per cm 1
Black band
0.007" ID (0.178 mm). About 0.25 µL per cm 1
Blue band
0.010" ID (0.254 mm). About 0.51 µL per cm 1
Yellow band
0.020" ID (0.508 mm). About 2.0 µL per cm 1
White band
0.030" ID (0.762 mm). About 4.6 µL per cm 1
Volume formula
Volume = π × (ID/2)² × length. I worked the µL per cm above from it 2

What goes wrong

  • Higher pressure than expected after re-plumbingCheck: A narrower ID somewhere. IDEX says the ID often has the biggest effect on pressure. 2

Sources

  1. IDEX Health & Science, Color Coding for 1/16" Stainless Steel Capillary Tubing (web page, saved 5 Nov 2026). colour table for IDEX precut 1/16" stainless steel capillary tubing.
  2. IDEX Health & Science, Fittings 101: A Practical Guide to Connections and Fittings in Your Laboratory (2025). What's the big ID?, p. 57; conversion formulas, p. 64.

New modules by email

All the modules are open now, and I keep adding parts and checks. Leave your email and I will send you what's new, including the Empower pages.

My HPLC newsletter too, about once a week. Unsubscribe any time. Privacy